Comparing Container Closure Integrity Test Methods-Performance of Headspace Carbon Dioxide Analysis versus Helium
Christian Proff1, Ken Victor2, Allison Alix Caudill2
1F. Hoffmann-La Roche Ltd, Global Device and Packaging Development, PTDZ, 4070 Basel, Switzerland; and christian.proff@roche.com.
PDA Journal of Pharmaceutical Science and Technology
|July 25, 2024
Summary
Comparing high-sensitivity package leak detection methods, this study found excellent correlation (R² ≥ 0.98) between laser-based gas headspace analysis and helium leakage testing for vials. Both methods proved accurate for package seal quality assessment.
Area of Science:
- Pharmaceutical Packaging
- Analytical Chemistry
- Quality Control
Background:
- Deterministic leak test methods, such as laser-based gas headspace analysis and helium leakage, offer high sensitivity for package integrity.
- No single package leak test method is universally applicable across all product-package systems.
- Understanding the comparative advantages and limitations of these sensitive techniques is crucial for effective package validation.
Purpose of the Study:
- To systematically compare the performance of laser-based gas headspace analysis and helium leakage testing.
- To evaluate the correlation between these two high-sensitivity leak detection methods.
- To assess the applicability of these techniques for package seal quality testing in pharmaceutical vials.
Main Methods:
- Utilized identical positive controls with microcapillaries for both measurement techniques.
- Gas headspace analysis measured carbon dioxide ingress rates under varying CO2 overpressures (0.5, 1.0, 2.0 bar).
- Helium leakage technique measured ingress rates using a standard helium leak detector with a 100% helium atmosphere.
Main Results:
- A strong correlation (minimum R² = 0.98) was observed between helium and carbon dioxide leakage rates for both ISO 2 R and ISO 10 R vials.
- Both testing setups met accuracy acceptance criteria when compared to calibrated standards.
- The study demonstrated the reliability of both methods for quantifying ingress rates.
Conclusions:
- Laser-based gas headspace analysis and helium leakage testing are highly correlated and accurate methods for package leak detection.
- These deterministic methods provide reliable data for package seal quality assessment.
- The findings support the use of either technique, depending on specific application needs, for ensuring pharmaceutical product integrity.