Related Experiment Video
Updated: Jun 23, 2026

Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Reliability of near-infrared spectroscopy in measuring muscle oxygenation during squat exercise
Juan Corral-Pérez1, Alberto Marín-Galindo1, Manuel Costilla1
1ExPhy Research Group, Department of Physical Education, Instituto de Investigación e Innovación Biomédica de Cádiz (INiBICA), Universidad de Cádiz, Spain.
Abstract:
Monitoring of changes in skeletal muscle oxygenation during exercise has increased in recent years. Tissue oxygenation, which is related to fatigue and muscle hypertrophy, is often measured using near-infrared spectroscopy (NIRS).
Objectives:
This study aimed to determine the test-retest reliability of a non-portable NIRS (NIRO200Nx) during the full-squat exercise and recovery in young healthy men.
Design:
Twenty-five male participants (21.8 ± 2.6 years) were recruited for this original research. Each participant completed an 8-repetition test with a load that elicited a velocity of 1 m·s-1. The test was conducted twice, with a 48-hour washout period between sessions.
Methods:
The NIRS measured the changes of oxygenated-Hemoglobin (O2Hb), deoxygenated-Hemoglobin (HHb) and Tissue Oxygenation Index (TOI) in both Vastus Lateralis and Vastus Medialis during rest, exercise, and recovery. Coefficient of Variation (CV), Standard Error Measurement (SEM) and Intraclass Correlation Coefficient (ICC) were used to evaluate the reliability of the data. Significance was set at p < 0.05.
Results:
The results indicated that TOI had good to acceptable absolute reliability (CVTOI = 2.7-10.2 %). A good relative relativity for the overall test was found for Vastus Medialis O2Hb (ICC = 0.851), HHb (ICC = 0.852), and TOI (ICC = 0.864), and Vastus Lateralis O2Hb (ICC = 0.898), HHb (ICC = 0.899), and TOI (ICC = 0.897).
Conclusions:
We conclude that NIRO200Nx is a reliable instrument for measuring muscle oxygen saturation through the TOI parameter in not-to-failure dynamic resistance exercises (1 set of 8 reps against ∼40 % 1 repetition maximum). Tissue oxygenation assessment could be a new way of individualizing exercise through dynamic resistance exercises.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024