Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey
Mateusz Kecik1,2, Martina Kropp1,2, Gabriele Thumann1,2
1Department of Ophthalmology, University of Geneva, 1205 Geneva, Switzerland.
Abstract:
Background and Objectives: The purpose of this study was to compare two commercially available specular microscopes (Tomey EM-4000 and Nidek CEM-530) in a real-life clinical setting in terms of intra- and interdevice variability. The study was conducted on all patients seen in a clinical practice specializing in anterior segment pathologies, regardless of the purpose of their visit. Materials and Methods: In total, 112 eyes of 56 patients (age 23-85 years old) were included in the study. Each eye was measured three times with each device (for a total of six measurements), and results for central corneal thickness (CCT) and corneal endothelial cell density (ECD) were recorded. The results were then evaluated with the D'Agostino-Pearson normality test and compared with a Wilcoxon signed-rank test, t-test, ANOVA or Mann-Whitney test for intra- and interdevice variability. Results: Both specular microscopes produced very reliable reproducible intradevice results: The Tomey EM-4000 measured an ECD of 2390 ± 49.57 cells/mm2 (mean ± standard error of mean); the range was 799-3010 cells/mm2. The determined CCT was 546 ± 5.104 µm (mean ± standard error of mean [SEM]); the range was 425-615 µm. The measurements with the Nidek CEM-530 revealed an ECD of 2417 ± 0.09 cells/mm2 (mean ± SEM); the range was 505-3461 cells/mm2 (mean ± SEM). The mean CCT detected was 546.3 ± 4.937 µm (mean ± SEM); the range was 431-621 µm. The interdevice differences were statistically significant for both parameters, ECD (p = 0.0175) and CCT (p = 0.0125) (p < 0.05). Conclusions: The Nidek CEM-530 and the Tomey EM-4000 both produced reliable and reproducible results in terms of ECD and CCT. The absolute measurements were statistically significantly different for CCT and ECD for both devices; the Nidek produces slightly higher values.
Related Concept Videos
Overview of Electron Microscopy
Transmission Electron Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...