Biomechanical Behavior of Injected Cement Spacers versus Traditional Cages in Low-Density Lumbar Spine under
Tibor Csakany1, Peter Varga2, Boyko Gueorguiev2
1National Center for Spinal Disorders, 1126 Budapest, Hungary.
Abstract:
Background and Objectives: Osteoporosis renders the use of traditional interbody cages potentially dangerous given the high risk of damage in the bone-implant interface. Instead, injected cement spacers can be applied as interbody devices; however, this technique has been mainly used in cervical spine surgery. This study aimed at investigating the biomechanical behavior of cement spacers versus traditional cages in lumbar spine surgery. Materials and Methods: Destructive monotonic axial compression testing was performed on 20 human cadaveric low-density lumbar segments from elderly donors (14 f/6 m, 70.3 ± 12.0 y) treated with either injected cement spacers (n = 10) or traditional cages (n = 10) without posterior instrumentation. Stiffness, failure load and displacement were compared. The effects of bone density, vertebral geometry and spacer contact area were evaluated. Results: Cement spacers demonstrated higher stiffness, significantly smaller displacement (p < 0.001) and a similar failure load compared to traditional cages. In the cage group, stiffness and failure load depended strongly on bone density and vertebral height, whereas failure displacement depended on vertebral anterior height. No such correlations were identified with cement spacers. Conclusions: Cement spacers used in lumbar interbody stabilization provided similar compression strength, significantly smaller failure displacement and a stiffer construct than traditional cages that provided benefits mainly for large and strong vertebrae. Cement stabilization was less sensitive to density and could be more beneficial also for segments with smaller and less dense vertebrae. In contrast to the injection of cement spacers, the optimal insertion of cages into the irregular intervertebral space is challenging and risks damaging bone. Further studies are required to corroborate these findings and the treatment selection thresholds.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024