A simple numerical simulation model can elucidate the key factors for designing a miniaturized ion trap mass
Xinming Huo1, Lin Gan1, Xinyue Ding2
1School of Biomedical Engineering, Shenzhen Campus of Sun Yat-sen University, Shenzhen, 518107, People's Republic of China; Key Laboratory of Sensing Technology and Biomedical Instruments of Guangdong Province, School of Biomedical Engineering, Sun Yat-sen University, Guangzhou, 510275, People's Republic of China.
Background:
Miniature ion trap mass spectrometer enables mass-to-charge ratio analysis of ions via quadrupole field in a low vacuum environment. It plays an important role in on-site detection due to its portability and specificity. In order to gain a deeper understanding of the analysis mechanism of miniature ion trap mass spectrometers, a quadrupole MS ion trajectory numerical simulation model (QITNS) is established in this paper for ions trajectory calculation under the action of quadrupole field, exciting field and neutral gas molecule collision. Compared with the existing methods, the model in this paper is simpler and more direct, which effectively explored the effects of dipole excitation and quadrupole excitation on ion manipulation under high background pressure.
Results:
The simulation results demonstrate that high RF amplitude, low auxiliary AC amplitude and quadrupole excitation can effectively improve the isolation resolution. Besides, it clarified the difference between the analysis mechanism of ion trap mass spectrometers under high background pressure (above 13.332 Pa) and absolute vacuum conditions. The relevant results are consistent with the conclusions of previous experiments and other theories, proving the applicability and accuracy of the proposed calculation model and solution method.
Significance:
This research bears the guiding significance for further understanding the mechanism of quadrupole mass spectrometry as well as designing and developing miniature mass spectrometers.
Related Concept Videos
Mass Analyzers: Common Types
Mass Spectrometers
Mass Analyzers: Overview
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...