A Feasibility Study of AI-Assisted Compressed Sensing in Prostate T2-Weighted Imaging
Liting Shen1, Hui Xu1, Qian Liao1
1Department of Radiology, Beijing Friendship Hospital, Capital Medical University, Beijing 100050, China (L.S., H.X., Q.L., Y.Y., Z.Y., L.W.).
Rationale And Objectives:
To evaluate the image quality and PI-RADS scoring performance of prostate T2-weighted imaging (T2WI) based on AI-assisted compressed sensing (ACS).
Materials And Methods:
In this prospective study, adult male urological outpatients or inpatients underwent prostate MRI, including T2WI, diffusion-weighted imaging and apparent diffusion coefficient maps. Three accelerated scanning protocols using parallel imaging (PI) and ACS: T2WIPI, T2WIACS1 and T2WIACS2 were evaluated through comparative analysis. Quantitative analysis included signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), slope profile, and edge rise distance (ERD). Image quality was qualitatively assessed using a five-point Likert scale (ranging from 1 = non-diagnostic to 5 = excellent). PI-RADS scores were determined for the largest or most suspicious lesions in each patient. The Friedman test and one-way ANOVA with post hoc tests were utilized for group comparisons, with statistical significance set at P < 0.05.
Results:
This study included 40 participants. Compared to PI, ACS reduced acquisition time by over 50%, significantly enhancing the CNR of sagittal and axial T2WI (P < 0.05), significantly improving the image quality of sagittal and axial T2WI (P < 0.05). No significant differences were observed in slope profile, ERD, and PI-RADS scores between groups (P > 0.05).
Conclusion:
ACS reduced prostate T2WI acquisition time by half while improving image quality without affecting PI-RADS scores.