Exploring single-shot propagation and speckle based phase recovery techniques for object thickness estimation by
Diego Rosich1,2, Margarita Chevalier1, Adrián Belarra1
1Complutense University of Madrid, Department of Radiology, Rehabilitation, and Physiotherapy, Faculty of Medicine, Madrid, Spain.
Purpose:
Propagation and speckle-based techniques allow reconstruction of the phase of an X-ray beam with a simple experimental setup. Furthermore, their implementation is feasible using low-coherence laboratory X-ray sources. We investigate different approaches to include X-ray polychromaticity for sample thickness recovery using such techniques.
Approach:
Single-shot Paganin (PT) and Arhatari (AT) propagation-based and speckle-based (ST) techniques are considered. The radiation beam polychromaticity is addressed using three different averaging approaches. The emission-detection process is considered for modulating the X-ray beam spectrum. Reconstructed thickness of three nylon-6 fibers with diameters in the millimeter-range, placed at various object-detector distances are analyzed. In addition, the thickness of an in-house made breast phantom is recovered by using multi-material Paganin's technique (MPT) and compared with micro-CT data.
Results:
The best quantitative result is obtained for the PT and ST combined with sample thickness averaging (TA) approach that involves individual thickness recovery for each X-ray spectral component and the smallest considered object-detector distance. The error in the recovered fiber diameters for both techniques is , despite the higher noise level in ST images. All cases provide estimates of fiber diameter ratios with an error of 3% with respect to the nominal diameter ratios. The breast phantom thickness difference between MPT-TA and micro-CT is about 10%.
Conclusions:
We demonstrate the single-shot PT-TA and ST-TA techniques feasibility for thickness recovery of millimeter-sized samples using polychromatic microfocus X-ray sources. The application of MPT-TA for thicker and multi-material samples is promising.