Related Experiment Video
Updated: Jun 18, 2025

Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Early Shear Failure of a 3-Peg Modified Dome Patellar Implant
Yifan V Mao1, Matthew V Dipane2, Erik N Zeegen2
1UCLA David Geffen School of Medicine, Los Angeles, CA, USA.
Abstract:
Patellar failure in total knee arthroplasty is a major source of complications postoperatively. Previous patellar failure reports commonly cited dissociations of modular and metal-backed patellar implants. However, mechanical breakage of monoblock all-polyethylene patellar implants is very rare. We present a case of an early shear failure of a 3-peg modified dome all-polyethylene patellar implant at 16 months. The patient underwent a revision procedure and at 1-year follow-up, the patient's patella and knee remained stable with no reported issues. Shear failure of polyethylene pegs requires excess cyclic shear stress imparted at the prosthetic-bone interface. Patellar implants with a cone design are more constrained and, if misaligned relative to the metallic trochlea, may impart excess shear force to the patella during flexion.

