Comparative accuracy of intraocular lens power calculation formulas when targeting myopia
Soonwon Yang1, Chanjoon Park2, Yong-Soo Byun2
1Department of Ophthalmology, Eunpyeong St. Mary's Hospital, The Catholic University of Korea, College of Medicine, Seoul, Republic of Korea.
Purpose:
This study aims to compare the accuracies of intraocular lens (IOL) power calculation formulas when targeting myopia versus emmetropia.
Methods:
A total of 450 patients were included, with 225 patients targeting emmetropia and 225 patients aiming for approximately -2.0 diopters of myopia. This retrospective analysis utilized data from a single eye of each patient, with preoperative biometric measurements obtained using the IOL Master 700. The study considered established formulas such as Haigis, Hoffer Q, Holladay 1, Holladay 2, and SRK/T, as well as modern formulas including Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF, Hoffer QST, Kane, Olsen, and PEARL-DGS. Statistical analyses, including Friedman test and post hoc analysis, were employed to compare the accuracy of each IOL power calculation formula between the two groups. Additionally, a multiple regression analysis was conducted to identify variables influencing the accuracy of intraocular lens power calculation formulas.
Results:
In targeting myopia, all IOL formulas tended to exhibit a greater refractive error compared to when targeting emmetropic eyes. Notably, the Haigis, SRK/T, and Holladay 2 formulas were found to be highly influenced by this trend, while the modern formulas were less affected.
Conclusion:
The accuracy of IOL power calculation formulas diminishes when targeting myopia in comparison to emmetropia. However, the modern formulas appear less susceptible to this trend. Consequently, when aiming for myopia, the use of the modern formulas is recommended for enhanced accuracy in IOL power calculation.
More Related Videos
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...