Objectives:
We sought to establish normal reference ranges (NRRs) for a novel TEG 6s cartridge (TEG 6s Citrated: K, KH, RTH, FFH [Global Hemostasis]) (Haemonetics Corporation, Boston, MA, US).
Methods:
Healthy volunteers (≥18 years of age) included in this single-arm study provided single samples of whole blood. Primary end points included TEG parameters in the citrated kaolin (CK), CK with heparinase (CKH), RapidTEG with heparinase (CRTH), and functional fibrinogen with heparinase (CFFH) assays.
Results:
Evaluable data were contributed by 164 volunteers (48.8% female; 62% White/Caucasian). The following NRRs were established: CK maximum amplitude (MA), 51.0 to 67.6 mm; CKH-MA, 51.8 to 67.9 mm; CRTH-MA, 53.0 to 68.9 mm; CFFH-MA, 15.3 to 34.4 mm; CK reaction time, 5.0 to 9.1 minutes; CKH reaction time, 4.9 to 9.4 minutes; CKH lysis 30 minutes after MA, 0% to 3.2%. Duplicate measurements demonstrated high reproducibility. CFFH-MA correlated with Clauss fibrinogen concentration (Pearson correlation coefficient, 0.74). Laboratory-based studies demonstrated maintenance of the relationship between CFFH-MA and fibrinogen up to 1344 mg/dL (hyperfibrinogenemic samples) and acceptability of heparin neutralization up to concentrations of low molecular weight and unfractionated heparin of 1.3 IU/mL and 5 IU/mL, respectively.
Conclusions:
This study established NRRs for the Global Hemostasis cartridge and serves as a proof of concept for the validity of results obtained using this cartridge.