Dual-Split CT to Simulate Multiple Radiation Doses From a Single Scan-Liver Lesion Detection Compared With
Damien Racine1, Tilo Niemann, Bence Nemeth
1From the Institute of Radiation Physics, University Hospital Lausanne (CHUV), University of Lausanne, Lausanne, Switzerland (D.R., L.G.M., A.V.); Department of Radiology, Kantonsspital Baden, Affiliated Hospital for Research and Teaching of the Faculty of Medicine of the University of Zurich, Baden, Switzerland (T.N., R.A.K.-H., A.E.); Department of Biomedical Imaging and Image-Guided Therapy, Division of Neuroradiology and Musculoskeletal Radiology, Medical University of Vienna, Vienna, Austria (B.N.); and Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland (H.A., T.F.).
Objectives:
The aim of this study was to evaluate the potential use of simulated radiation doses from a dual-split CT scan for dose optimization by comparing their lesion detectability to dose-matched single-energy CT acquisitions at different radiation dose levels using a mathematical model observer.
Materials And Methods:
An anthropomorphic abdominal phantom with liver lesions (5-10 mm, both hyperattenuating and hypoattenuating) was imaged using a third-generation dual-source CT in single-energy dual-source mode at 100 kVp and 3 radiation doses (5, 2.5, 1.25 mGy). The tube current was 67% for tube A and 33% for tube B. For each dose, 5 simulated radiation doses (100%, 67%, 55%, 45%, 39%, and 33%) were generated through linear image blending. The phantom was also imaged using traditional single-source single-energy mode at equivalent doses. Each setup was repeated 10 times. Image noise texture was evaluated by the average spatial frequency (f av ) of the noise power spectrum. Liver lesion detection was measured by the area under the receiver operating curve (AUC), using a channelized Hotelling model observer with 10 dense Gaussian channels.
Results:
F av decreased at lower radiation doses and differed between simulated and single-energy images (eg, 0.16 mm -1 vs 0.14 mm -1 for simulated and single-energy images at 1.25 mGy), indicating slightly blotchier noise texture for dual-split CT. For hyperattenuating lesions, the mean AUC ranged between 0.92-0.99, 0.81-0.96, and 0.68-0.89 for single-energy, and between 0.91-0.99, 0.78-0.91, and 0.70-0.85 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. For hypoattenuating lesions, the AUC ranged between 0.90-0.98, 0.75-0.93, and 0.69-0.86 for the single-energy, and between 0.92-0.99, 0.76-0.87, and 0.67-0.81 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. AUC values were similar between both modes at 5 mGy, and slightly lower, albeit not significantly, for the dual-split mode at 2.5 and 1.25 mGy.
Conclusions:
Lesion detectability was comparable between multiple simulated radiation doses from a dual-split CT scan and dose-matched single-energy CT. Noise texture was slightly blotchier in the simulated images. Simulated doses using dual-split CT can be used to assess the impact of radiation dose reduction on lesion detectability without the need for repeated patient scans.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET