Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Improved image quality with deep learning reconstruction - a study on a semi-anthropomorphic upper-abdomen phantom
Tormund Njølstad1,2, Anselm Schulz1, Kristin Jensen3
1Department of Radiology and Nuclear Medicine, Oslo University Hospital Ullevål, Oslo 0450, Norway.
Purpose:
To assess image quality of a deep learning reconstruction (DLR) algorithm across dose levels using a semi-anthropomorphic upper-abdominal phantom, and compare with filtered back projection (FBP) and hybrid iterative reconstruction (IR).
Material And Methods:
CT scans obtained at five dose levels (CTDIvol 5, 10, 15, 20 and 25 mGy) were reconstructed with FBP, hybrid IR (IR50, IR70 and IR90) and DLR of low (DLL), medium (DLM) and high strength (DLH) in 0.625 mm and 2.5 mm slices. CT number, homogeneity, noise, contrast, contrast-to-noise ratio (CNR), noise texture deviation (NTD; a measure of IR-specific artifacts), noise power spectrum (NPS) and task-based transfer function (TTF) were compared between reconstruction algorithms.
Results:
CT numbers were highly consistent across reconstruction algorithms. Image noise was significantly reduced with higher levels of DLR. Noise texture (NPS and NTD) was with DLR maintained at comparable levels to FBP, contrary to increasing levels of hybrid IR. Images reconstructed with DLR of low and high strength in 0.625 mm slices showed similar noise characteristics to 2.5 mm slice FBP and IR50, respectively. Dose-reduction potential based on image noise with IR50 as reference was estimated to 35% for DLM and 74% for DLH.
Conclusions:
The novel DLR algorithm demonstrates robust noise reduction with maintained noise texture characteristics despite higher algorithm strength, and may have overcome important limitations of IR. There may be potential for dose reduction and additional benefit from thin-slice reconstruction.