Related Experiment Video
Updated: Jun 18, 2025

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
The temperature profile in elbow arthroscopy using radiofrequency ablation.
Felix Krane1, Kilian Wegmann2, Stephanie Kahmann3
1Faculty of Medicine and University Hospital, Center for Orthopedic, Trauma and Plastic Surgery, University of Cologne, Kerpener Str. 62, 50937, Cologne, Germany. felix.krane1@uk-koeln.de.
Radiofrequency ablation (RFA) in arthroscopic elbow surgery is safe with irrigation. Continuous use without irrigation should be limited to 3 seconds to prevent dangerous temperature increases near the ulnar nerve.
Area of Science:
- Orthopedic Surgery
- Minimally Invasive Procedures
- Biomedical Engineering
Background:
- Arthroscopic surgery is increasingly utilized for its minimally invasive benefits.
- Elbow arthroscopy presents unique challenges and potential risks.
- Investigating thermal effects during radiofrequency ablation is crucial for patient safety.
Purpose of the Study:
- To assess temperature development in the elbow joint during radiofrequency ablation in arthroscopic surgery.
- To evaluate the safety of radiofrequency ablation concerning thermal spread.
- To determine safe usage parameters for radiofrequency ablation in elbow arthroscopy.
Main Methods:
- Cadaveric elbow models (n=7) underwent arthroscopic procedures with posterolateral, transtricipital, and anterolateral approaches.
- Bipolar radiofrequency ablation (bRFA) was applied at specific locations with and without irrigation.
- Temperature probes monitored joint temperatures, particularly near the olecranon fossa and ulnar nerve.
Main Results:
- Radiofrequency ablation without irrigation caused rapid intra-articular temperature increases within 5 seconds.
- Significant temperature rise was observed in the olecranon fossa without irrigation (p=0.0052).
- With continuous irrigation, temperatures near the ulnar nerve remained below 41°C within 30 seconds (p=0.0747).
Conclusions:
- Radiofrequency ablation is safe for elbow arthroscopy when utilized with continuous irrigation.
- Unirrigated radiofrequency ablation should be limited to 3-second applications.
- Thermal effects of radiofrequency ablation are unlikely to adversely impact the ulnar nerve despite its proximity.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...