Efficient use of penile Doppler ultrasound for investigating men with erectile dysfunction
Jose M Flores1, Michael West1, John P Mulhall1
1Sexual and Reproductive Medicine Program, Urology Service, Memorial Sloan Kettering Cancer Center, New York, NY 10065, United States.
Background:
Vasculogenic erectile dysfunction is the most common type of erectile dysfunction, and penile Doppler ultrasound (PDUS) is a useful tool to assess erectile hemodynamics in the clinician's effort to discuss prognosis and management strategies with the patient.
Aim:
We herein describe the PDUS protocol used at our center, including indications, technique, and data interpretation.
Methods:
We describe our institutional experience with PDUS and discuss it in the context of a contemporary review of the literature for this investigation.
Outcome:
Our institutional PDUS protocol.
Results:
To perform PDUS properly, adequate training, equipment, setting, technique, and interpretation are critical. The accuracy of PDUS is entirely predicated on achieving complete cavernosal smooth muscle relaxation. A redosing protocol optimizes the reliability and reproducibility of the hemodynamic data acquired during PDUS. A rigidity-based assessment is performed, and patients are scanned according to the erection rigidity achieved (full hardness) or by administration of maximum dose of the vasoactive agent. Peak systolic velocity is considered a measure of arterial inflow (normal, >30 cm/s), while end diastolic velocity evaluates the veno-occlusive mechanism (normal, <5 cm/s). After the procedure, the patient is evaluated to confirm detumescence. If the patient has a persistent penetration rigidity erection, intracavernosal phenylephrine is administered; however, if detumescence is not achieved with intracavernosal phenylephrine injections alone, corporal aspiration is potentially performed.
Conclusion:
PDUS is a valuable minimally invasive tool for erectile hemodynamics assessment and an accurate assessment of such, provided that complete cavernosal smooth muscle relaxation is achieved.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...