Validation of a Novel Method for Noninvasive Mixed Venous Oxygen Saturation Monitoring in Anesthetized Children
Anders Svedmyr1,2, Kristoffer Steiner3,4, Andreas Andersson1,2
1From the Department of Physiology and Pharmacology (FYFA), C3, Eriksson I Lars, PA Lönnqvist Group, Section of Anesthesiology and Intensive Care, Karolinska Institute, Stockholm, Sweden.
Background:
Mixed venous oxygen saturation (SvO 2 ) is a critical variable in the assessment of oxygen supply and demand but is rarely used in children due to the invasive nature of pulmonary artery catheters. The aim of this prospective, observational study was to investigate the accuracy of noninvasively measured SvO 2 acquired by the novel capnodynamic method, based on differential Fick equation (Capno-SvO 2 ), against gold standard CO-oximetry.
Methods:
Capno-SvO 2 was compared to SvO 2 measured by pulmonary artery blood gas CO-oximetry in children undergoing cardiac catheter interventions and subjected to moderate hemodynamic challenges. Bland-Altman analysis was used to describe the agreement of absolute values between CO-oximetry and Capno-SvO 2 , and a concordance rate was calculated to evaluate the ability of Capno-SvO 2 to track change.
Results:
Twenty-five procedures were included in the study. Capno-SvO 2 showed a bias toward CO-oximetry of +3 percentage points; upper and lower limits of agreement were +11 percentage points (95% confidence interval [CI], 9-14) and -5 percentage points (95% CI, -8 to -3), respectively. The concordance rate was 92% (95% CI, 89-96).
Conclusions:
In conclusion, this first clinical application of a novel concept for noninvasive SvO 2 monitoring without the need for a pulmonary artery catheter indicates that Capno-SvO 2 generates absolute values and trending capacity in close agreement with the gold standard reference method.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...