Technical note: Phantom-based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy
Claas Wessels1, Adam Strzelecki1, Mathieu Plamondon1
1Varian Medical Systems Imaging Laboratory, Daettwil, Switzerland.
Background:
High-quality 3D-anatomy of the day is needed for treatment plan adaptation in radiotherapy. For online x-ray-based CBCT workflows, one approach is to create a synthetic CT or to utilize a fan-beam CT with corresponding registrations. The former potentially introduces uncertainties in the dose calculation if deformable image registration is used. The latter can introduce burden and complexity to the process, the facility, and the patient.
Purpose:
Using the CBCT of the day, acquired on the treatment device, for direct dose calculation and plan adaptation can overcome these limitations. This study aims to assess the accuracy of the calculated dose on the CBCT scans acquired on a Halcyon linear accelerator equipped with HyperSight.
Methods:
HyperSight's new CBCT reconstruction algorithm includes improvements in scatter correction, HU calibration of the imager, and beam shape adaptation. Furthermore, HyperSight introduced a new x-ray detector. To show the effect of the implemented improvements, gamma comparisons of 2%/2 mm, 2%/1 mm, and 1%/1 mm were made between the dose distribution in phantoms calculated on the CBCT reconstructions and the simulation CT scans, considering this the standard of care. The resulting gamma passing rates were compared to those obtained with the Halcyon 3.0 reconstruction and hardware without HyperSight's technologies. Various anatomical phantoms for dosimetric evaluations on brain, head and neck, lung, breast, and prostate cases have been used in this study.
Results:
The overall results demonstrated that HyperSight outperformed the Halcyon 3.0 version. Based on the gamma analysis, the calculated dose using HyperSight was closer to the CT scan-based doses than the calculated dose using iCBCT Halcyon 3.0 for most cases. Over all plans and gamma criteria, Halcyon 3.0 achieved an average passing rate of 92.9%, whereas HyperSight achieved 98.1%.
Conclusion:
Using HyperSight CBCT images for direct dose calculation, for example, in (online) plan adaptation, seems feasible for the investigated cases.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography