Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Dose rate correction for the novel 2D diode array MapCHECK 3
Mengyang Li1,2, Yuan Tian1,2, Linyi Shen1,2
1Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences, Beijing, China.
Purpose:
To investigate the dose rate dependence of MapCHECK3 and its influence on measurement accuracy, as well as the effect of dose rate correction.
Materials And Methods:
The average and instantaneous dose rate dependence of MapCHECK2 and MapCHECK3 were studied. The accuracy of measurements was investigated where the dose rate differed significantly between dose calibration of the MapCHECK and the measurement. Measurements investigated include: the central axis dose for different fields at different depths, off-axis doses outside the field, and off-axis doses along the wedge direction. Measurements using an ion chamber were taken as the reference. Exponential functions were fit to account for average and instantaneous dose rate dependence for MapCHECK3 and used for dose rate correction. The effect of the dose rate correction was studied by comparing the differences between the measurements for MapCHECK (with and without the correction) and the reference.
Results:
The maximum dose rate dependence of MapCHECK3 is greater than 2.5%. If the dose calibration factor derived from a 10 × 10 cm2 open field at 10 cm depth was used for measurements, the average differences in central diode dose were 0.8% ± 1.0% and 1.0% ± 0.8% for the studied field sizes and measurement depths, respectively. The introduction of wedge would not only induce -1.8% ± 1.3% difference in central diode dose, but also overestimate the effective wedge angle. After the instantaneous dose rate correction, above differences can be changed to 1.9% ± 8.1%, 0.2% ± 0.1%, and 0.0% ± 0.9%. The pass rate can be improved from 98.4% to 98.8%, 98.3%-100.0%, and 96.3%-100.0%, respectively.
Conclusion:
Compared with MapCHECK2 (SunPoint1 diodes), the more pronounced dose rate dependence of MapCHECK3 (SunPoint2 diodes) should be carefully considered. To ensure highly accurate measurement, it is suggested to perform the dose calibration at the same condition where measurement will be performed. Otherwise, the dose rate correction should be applied.