Background:
Left ventricular assist device (LVAD) malposition has been linked to hemocompatibility-related adverse events (HRAEs). This study aimed to identify preoperative anatomical landmarks and postoperative pump position, associated with HRAEs during LVAD support.
Methods:
Pre- and postoperative chest X-ray measures (≤14 days pre-implantation, first postoperative standing, 6, 12, 18, and 24 months post-implantation) were analyzed for their association with HRAEs over 24 months in 33 HeartMate 3 (HM3) patients (15.2% female, age 66 (9.5) years).
Results:
HM3 patients with any HRAE showed significantly lower preoperative distances between left ventricle and thoracic outline (dLVT) (25.3 ± 10.2 mm vs. 40.3 ± 15.5 mm, p = 0.004). A ROC-derived cutoff dLVT ≤ 29.2 mm provided 85.7% sensitivity and 72.2% specificity predicting any HRAE during HM3 support (76.2% (>29.2 mm) vs. 16.7% (≤29.2 mm) freedom from HRAE, p < 0.001) and significant differences in cardiothoracic ratio (0.58 ± 0.04 vs. 0.62 ± 0.04, p = 0.045). Postoperative X-rays indicated lower pump depths in patients with ischemic strokes (9.1 ± 16.2 mm vs. 38.0 ± 18.5 mm, p = 0.007), reduced freedom from any neurological event (pump depth ≤ 28.7 mm: 45.5% vs. 94.1%, p = 0.004), and a significant correlation between pump depth and inflow cannula angle (r = 0.66, p < 0.001). Longitudinal changes were observed in heart-pump width (F(4,60) = 5.61, p < 0.001).
Conclusion:
Preoperative X-ray markers are associated with postoperative HRAE occurrence. Applying this knowledge in clinical practice may enhance risk stratification, guide therapy optimization, and improve HM3 recipient management.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Anatomical Positions
The body is upright, facing forward, and standing erect.
The feet are parallel and flat on the floor.
The arms are hanging by the...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...