Extent of Radiolytic Damage from Liquid Cell TEM Experiments on Metal-Organic Frameworks via Post-Mortem 4D-STEM
Karthik Gnanasekaran1,2, Nathan D Rosenmann3, Roberto Dos Reis4,5
1Materials and Structural Analysis, Thermo Fisher Scientific Inc., 5350 NE Dawson Creek Drive Hillsboro, Oregon 97124, United States.
Abstract:
We report a systematic analysis of electron beam damage of the zeolitic imidazolate framework (ZIF-8) during liquid cell transmission electron microscopy (LCTEM). Our analysis reveals ZIF-8 morphology is strongly affected by solvent used (water vs dimethylformamide), electron flux applied, and imaging mode (i.e., TEM vs STEM), while ZIF-8 crystallinity is primarily affected by accumulated electron fluence. Our observations indicate that the stability of ZIF-8 morphology is higher in dimethylformamide (DMF) than in water. However, in situ electron diffraction indicates that ZIF-8 nanocrystals lose crystallinity at critical fluence of ∼80 e-Å-2 independent of the presence of solvent. Furthermore, 4D-STEM analysis as a post-mortem method reveals the extent of electron beam damage beyond the imaging area and indicates that radiolytic reactions are more pronounced in TEM mode than in STEM mode. These results illustrate the significance of radiolysis occurring while imaging ZIF-8 and present a workflow for assessing damage in LCTEM experiments.