Related Experiment Video
Updated: Jun 17, 2025

Disentangling High Strength Copolymer Aramid Fibers to Enable the Determination of Their Mechanical Properties
Published on: September 1, 2018
A multi-mode stretching apparatus for in situ synchrotron x-ray scattering of polymer materials
Bo Yang1,2, Zhen Yuan3, Xia Dong1,2
1Beijing National Laboratory for Molecular Sciences, CAS Key Laboratory of Engineering Plastics, Institute of Chemistry, Chinese Academy of Sciences, Beijing 100190, China.
Abstract:
A compact and versatile tensile apparatus for polymer materials is designed and fabricated. Three distinct stretching modes are developed: constant speed, cyclic, and sinusoidal, with adjustable speeds ranging from 0.001 to 120 mm/s. To capture the true strain of the central region, a high-speed camera has been integrated into the apparatus. The temperature of the sample chamber is controlled by flowing air, enabling a homogeneous temperature in the range of RT ∼200 °C. The apparatus is particularly suitable for a synchrotron beamline. The structural evolution of natural rubber during sinusoidal stretching is investigated by in situ wide-angle x-ray scattering. Scattering patterns, force, clamp position, and sample images are saved simultaneously during stretching. Notably, the results reveal a sinusoidal variation in the crystallinity of crosslinked natural rubber when a sinusoidal strain was applied to the sample. The integration of advanced measurement techniques and controlled experimental conditions ensures the acquisition of reliable and accurate data, providing valuable insights into the structural evolution of materials under dynamic deformation conditions.