Pediatric Advanced Life Support Tape for Indian Children

Vaman Khadilkar1, Sagar Lad2, Shruti Mondkar2

  • 1Pediatric Endocrinology Unit, Hirabai Cowasji Jehangir Medical Research Institute, Pune, India. Correspondence to: Dr. Vaman Khadilkar, Pediatric Endocrinology Unit, Jehangir Hospital, Pune, and Department of Health Sciences, Savitribai Phule Pune University, Pune, Maharashtra, India. vamankhadilkar@gmail.com.

Indian Pediatrics
|August 8, 2024
PubMed

Insights

A new advanced life support (ALS) tape, tailored for Indian children using local data, offers more accurate weight estimation than the Broselow tape. This improved accuracy can help reduce errors in pediatric emergency care.

Area of Science:

  • Pediatric Emergency Medicine
  • Biomedical Engineering
  • Growth Monitoring

Background:

  • Accurate weight estimation is critical for pediatric emergency care, influencing medication dosages and equipment selection.
  • Existing weight estimation tools, like the Broselow tape, may have limitations in diverse populations.
  • A need exists for a weight estimation tape specifically adapted to the anthropometric data of Indian children.

Purpose of the Study:

  • To develop and validate a novel advanced life support (ALS) tape using Indian multicenter height/length and weight data.
  • To ensure accurate weight estimation from recumbent length for pediatric patients in India.
  • To compare the accuracy of the new ALS tape against the established Broselow tape.

Main Methods:

  • Designed a new ALS tape by aligning median weights with median heights/lengths from recent Indian growth data.
  • Maintained the color-coding system consistent with the Broselow tape for ease of transition.
  • Validated the weight estimation accuracy of the new tape against the Broselow tape in a tertiary care hospital setting.

Main Results:

  • The new Indianized ALS tape demonstrated higher median height correlations for each color band compared to the Broselow tape.
  • A significant proportion of children, across various color bands, were assigned a higher Broselow color band, indicating potential weight overestimation by the Broselow tape.
  • The new ALS tape exhibited minimal percentage differences in estimated weight versus actual weight (-0.5% for under-5 years, 0.2% for older children).

Conclusions:

  • The developed Indianized ALS tape provides more accurate weight estimations for Indian children.
  • Implementation of this new ALS tape has the potential to decrease errors in calculating emergency medications, fluid volumes, and equipment sizes.
  • Further validation studies are recommended for the new ALS tape within Indian pediatric emergency departments.
Abstract