An impedimetric approach for determination of ammonium using silver/poly-1-aminoanthraquinone/carbon paste electrode
Mahmoud Fatehy Altahan1, Amr Mohamed Beltagi2, Magdi Abdel-Azzem3
1Central Laboratory for Environmental Quality Monitoring, National Water Research Center, El-Qanater El-Khairia, 13621, Egypt. mahmoud_abdalqader@nwrc.gov.eg.
Abstract:
Here, we present the electrochemical determination of ammonium in water samples, emphasizing the importance of accurate and precise assessment of its concentration. The modified electrode used in this study was fabricated through the anodic polymerization of 1-aminoanthraquinone (1-AAQ) and deposition of silver particles into a carbon paste electrode. The fabrication process involved cyclic voltammetry in a 0.1 M HCl solution, followed by the application of a potential of 0.2 V for 75 s. The resulting Ag/poly-1-AAQ/CPE exhibited remarkable electrochemical properties, as confirmed by scanning electron spectroscopy (SEM), energy-dispersive X-ray analysis (EDX), and elemental mapping. The successful deposition of silver at percentages of 12.07% on Ag/CPE and 0.75% on Ag/poly-1-AAQ/CPE was observed. The Ag/poly-1-AAQ/CPE was employed for impedimetric determination of ammonium in a solution of 0.1 M Na2SO4. The charge transfer resistance) output from the fitting of the experimental impedimetric data of ammonium determination exhibited good linearity over a concentration range of 5 µM to 200 µM NH4+, with a detection limit of 3.3 µM NH4+. The precision of the modified electrode over ten replicate measurements were conducted at three concentration levels (a low of 5 µM NH4+, a medium of 50 µM NH4+, and a high of 200 µM NH4+). The obtained relative standard deviation (RSD) values of 18%, 12% and 7%, respectively, indicating good precision.
Related Concept Videos
Amperometry: Overview
Titration of a Weak Base with a Strong Acid
Using the ICE table and substituting the Kb value, we calculate the initial pH of 50 mL of 0.1 M ammonia to be 11.11. Addition of 25 mL of 0.1 M hydrochloric acid to this solution of ammonia results in a buffer with an equal concentration of ammonia and ammonium ions. The pH of this buffer can be calculated by substituting these...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Titration of a Weak Acid with a Weak Base
As a result, there is no simple...
EDTA: Indirect and Alkalimetric Titration
Precipitation Titration: Overview
A precipitation titration curve demonstrates the change in concentration of the titrant or analyte upon adding the...