Evaluation of the Minimum Segment Width and Fluence Smoothing Tools for Intensity-modulated Techniques in Monaco
Sara Jiménez-Puertas1, Andrea González Rodríguez1, Sergio Lozares Cordero1
1Department of Medical Physics and Radiation Protection, University Hospital Miguel Servet, P.º de Isabel la Católica, 1-3, 50009 Zaragoza, Spain.
Purpose:
This study aims to minimize monitor units (MUs) of intensity-modulated treatments in the Monaco treatment planning system while preserving plan quality by optimizing the "Minimum Segment Width" (MSW) and "Fluence Smoothing" parameters.
Materials And Methods:
We retrospectively analyzed 30 prostate, 30 gynecological, 15 breast cancer, 10 head and neck tumor, 11 radiosurgery, and 10 hypo-fractionated plans. Original prostate plans employed "Fluence Smoothing" = Off and were reoptimized with Low, Medium, and High settings. The remaining pathologies initially used MSW = 0.5 cm and were reoptimized with MSW = 1.0 cm. Plan quality, including total MU, delivery time, and dosimetric constraints, was statistically analyzed with a paired t-test.
Results:
Prostate plans exhibited the highest MU variation when changing "Fluence Smoothing" from Off to High (average ΔMU = -5.1%; P < 0.001). However, a High setting may increase overall MU when MSW = 0.5 cm. Gynecological plans changed substantially when MSW increased from 0.5 cm to 1.0 cm (average ΔMU = -29%; P < 0.001). Organs at risk sparing and planning target volumes remained within 1.2% differences. Replanning other pathologies with MSW = 1.0 cm affected breast and head and neck tumor plans (average ΔMU = -168.38, average Δt = -11.74 s, and average ΔMU = -256.56, average Δt = -15.05 s, respectively; all with P < 0.004). Radiosurgery and hypofractioned highly modulated plans did not yield statistically significant results.
Conclusions:
In breast, pelvis, head and neck, and prostate plans, starting with MSW = 1.0 cm optimally reduces MU and treatment time without compromising plan quality. MSW has a greater impact on MU than the "Fluence Smoothing" parameter. Plans with high modulation might present divergent behavior, requiring a case-specific analysis with MSW values higher than 0.5 cm.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Stress Concentrations in Circular Shafts
Design of Transmission Shafts - Stress Analysis
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Typical Model Studies