Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
Published on: January 7, 2019
Comparison of human observer impression of X-ray fluoroscopy and angiography image quality with technical changes to
Jelena M Mihailovic1, Yoshihisa Kanaji2, Daniel Miller2
1Mayo Clinic, Department of Radiology, Rochester, Minnesota, United States.
Purpose:
Spatio-temporal variability in clinical fluoroscopy and cine angiography images combined with nonlinear image processing prevents the application of traditional image quality measurements in the cardiac catheterization laboratory. We aimed to develop and validate methods to measure human observer impressions of the image quality.
Approach:
Multi-frame images of the thorax of a euthanized pig were acquired to provide an anatomical background. The detector dose was varied from 6 to 200 nGy (increments 2×), and 0.6 and 1.0 mm focal spots were used. Two coronary stents with/without 0.5 mm separation and a synthetic right coronary artery (RCA) with hemispherical defects were embedded into the background images as test objects. The quantitative observer ( ) performance was measured using a two-alternating forced-choice test of whether stents were separated and by a count of visible right coronary artery defects. Qualitative impressions of noise, spatial resolution, and overall image quality were measured using a visual analog scale (VAS). A paired -test and multinomial logistic regression model were used to identify statistically significant factors affecting the observer's impression image quality.
Results:
The proportion of correct detection of stent separation and the number of reported right coronary artery defects changed significantly with detector dose increment in the 6 to 100 nGy ( ). Although a trend favored the 0.6 versus 1.0 mm focal spot for these quantitative assessments, this was insignificant. Visual analog scale measurements changed significantly with detector dose increments in the range of 24 to 100 nGy and focal spot size ( ). The application of multinomial logistic regression analysis to observer VAS scores demonstrated sensitivity matching of the paired -test applied to quantitative observer performance measurements.
Conclusions:
Both quantitative and qualitative measurements of observer impression of the image quality were sensitive to image quality changes associated with changing the detector dose and focal spot size. These findings encourage future work that uses qualitative image quality measurements to assess clinical fluoroscopy and angiography image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT