Variations in head tilt during the acquisition of cone beam computed tomography scans and their effects on effective
Arthur X M Mancini1, Gabriela A Carmozini2, Thiago M Inácio2
1Department of Dental Materials and Prosthodontics, School of Dentistry of Ribeirao Preto, University of Sao Paulo, Ribeirao Preto, Sao Paulo, 14040-904, Brazil.
Objectives:
This study evaluated the impact of variations in anteroposterior and lateral tilts of patients head on radiation-weighted doses to organs/tissues and effective doses using 3 different cone beam computed tomography (CBCT) machines.
Methods:
An anthropomorphic phantom was used to estimate radiation doses in 3 CBCT machines (OP300, Eagle X 3D, and Eagle Edge). Thermoluminescent dosimeters were placed in regions corresponding to pre-stablished organ/tissues. CBCT examinations from the posterior mandible and anterior maxilla regions were acquired, with 3 different anteroposterior angulations (0°, 30°, and 45°), and from the posterior mandible in 3 different lateral angulations (0°, 20° to the left, and 20° to the right side). Radiation-weighted doses for each organ/tissue and effective doses were calculated for each machine and angulation.
Results:
For the posterior mandible acquisitions, anteroposterior angulations of the head at 30° and 45° yielded a reduction in effective doses in all 3 devices. A 20° tilt to the right side resulted in lower doses than to the left [same side as the field of view (FOV)]. For the anterior maxilla, increased anteroposterior angulation was associated with reduction in effective dose in 2 devices.
Conclusions:
Effective doses are lower when small FOV CBCT exams of the posterior mandible and anterior maxilla are acquired with increased anteroposterior head angulation at 30° and 45°. For FOV in the posterior mandible, a 20° lateral tilt towards the side opposite to the FOV also yields lower effective doses. The main contribution to these dose reductions is the decrease in dose to salivary glands.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT