State-of-the-Art Quantification of Liver Iron With MRI-Vendor Implementation and Available Tools

Aaryani Tipirneni-Sajja1,2, Utsav Shrestha1,2, Juan Esparza1

  • 1Department of Biomedical Engineering, The University of Memphis, Memphis, Tennessee, USA.

Insights

Magnetic Resonance Imaging (MRI) estimates liver iron concentration (LIC) to manage iron overload. R2*-MRI is emerging as the most accurate and reproducible method for LIC estimation, aiming for clinical standardization.

Area of Science:

  • Radiology
  • Medical Imaging
  • Biophysics

Background:

  • Magnetic Resonance Imaging (MRI) is crucial for estimating liver iron concentration (LIC) in patients with iron overload.
  • Standardization of MRI-based LIC quantification is limited by multiple existing techniques.

Purpose of the Study:

  • To review established and emerging MRI techniques for LIC estimation.
  • To discuss protocols, calibrations, software, and clinical implications for MRI-based LIC quantification.

Main Methods:

  • Review of signal intensity ratio (SIR), R2, and R2* relaxometry methods for LIC estimation at 1.5T and 3T.
  • Discussion of acquisition protocols, analysis, calibration, vendor implementations, and software.

Main Results:

  • Both R2 and R2* relaxometry can accurately estimate LIC with validated protocols.
  • R2*-MRI is increasingly endorsed as the most accurate and reproducible method over the Ferriscan R2 method.
  • Emerging R2*-MRI techniques show promise with improved motion compensation and dynamic range.

Conclusions:

  • R2*-MRI is positioned to become the standard for LIC quantification.
  • Standardized MRI-based LIC measurements will improve clinical decision-making in iron overload management.