Related Experiment Video
Updated: Jul 23, 2026

Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
Evaluation of lingual oscillometric blood pressure measurement in anaesthetized pigs
Charles Makepeace1, Paul Macfarlane1, Julia Deutsch1
1Langford Vets and Translational Biomedical Research Centre, University of Bristol, Langford, Bristol, UK.
Objective:
To evaluate the agreement between measurements of invasive blood pressure (IBP) and oscillometric blood pressure from the tongue (OBPton) using a multiparameter monitor.
Study Design:
Unblinded, prospective, experimental study.
Animals:
A total of 12 female Large White crossbreed pigs.
Methods:
Pigs undergoing experimental procedures that required arterial cannula placement were recruited. A blood pressure cuff with the closest width to 40% of the circumference of the tongue was placed rostral to the lingual frenulum. Systolic, mean and diastolic IBP and OBPton were measured simultaneously at 5 minute intervals. Agreement between paired measurements was examined using Bland-Altman analysis. Mean bias, precision (standard deviation of mean bias), 95% limits of agreement, correlation coefficients and percentage of measurements within 10 and 20 mmHg of IBP were calculated.
Results:
The total numbers of paired measurements recorded were 124, 126 and 124 for systolic, mean and diastolic blood pressures, respectively. The mean bias, precision and 95% limits of agreement for systolic OBPton were 11.5, 11.5 (-11.1 to 34.2), for mean OBPton 5.6, 5.7 (-5.7 to 16.8) and for diastolic OBPton 7.6, 10.1 (-12.1 to 27.4) mmHg. Correlation coefficients were greater than 0.9 for mean OBPton only. More than 50% of measurements were within 10 mmHg of IBP and 80% of measurements were within 20 mmHg of IBP for mean and diastolic OBPton only.
Conclusions And Clinical Relevance:
The tongue as a cuff site for oscillometric blood pressure measurement is a useful site for measuring mean arterial, but not systolic or diastolic blood pressure in anaesthetized Large White crossbreed pigs. This technique fulfils the American College of Veterinary Internal Medicine criteria for measuring mean arterial pressure but not systolic or diastolic arterial pressure.