The impact of hyperbaric oxygen treatment for cardiovascular implantable electronic devices
Kentaro Goto1, Shinsuke Miyazaki1, Takuya Oyaizu2
1Department of Cardiovascular Medicine Tokyo Medical and Dental University Hospital Tokyo Japan.
Introduction:
The safety of hyperbaric oxygen treatment (HBO2) in patients with cardiovascular implanted electronic devices (CIED) remains unclear.
Methods:
We conducted a retrospective analysis of seven CIED patients (median age 79 [73-83] years, five males [71.4%]), including five with pacemakers and two with implantable cardioverter defibrillators (ICD), who underwent HBO2 between June 2013 and April 2023. During the initial session, electrocardiogram monitoring was conducted, and CIED checks were performed before and after the treatment. In addition, the medical records were scrutinized to identify any abnormal CIED operations.
Results:
All seven CIED patients underwent HBO2 within the safety pressure range specified by the CIED manufacturers or general pressure test by the International Organization for Standardization (2.5 [2.5-2.5] atmosphere absolute × 18 [5-20] sessions). When comparing the CIED parameters before and after HBO2, no significant changes were observed in the waveform amplitudes, pacing thresholds, lead impedance of the atrial and ventricular leads, or battery levels. All seven patients, including two with the rate response function activated, exhibited no significant changes in the pacing rate or pacing failure. Two ICD patients did not deactivate the therapy, including the defibrillation; however, they did not experience any arrhythmia or inappropriate ICD therapy during the HBO2.
Conclusion:
CIED patients who underwent HBO2 within the safety pressure range exhibited no significant changes in the parameters immediately after the HBO2 and had no observable abnormal CIED operations during the treatment. The safety of defibrillation by an ICD during HBO2 should be clarified.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Heart Failure VI: Adjunct Therapies