A novel chemiluminescence sensor for alpha-fetoprotein detection based on an aptamer-luminol modified magnetic
Juntao Liu1, Xueying Wang2, Yuanling Sun2
1Henan University of Animal Husbandry and Economy, Zhengzhou 450046, China.
Abstract:
Herein, an aptamer-luminol modified magnetic graphene oxide and copper-based MOF composite was prepared and used to build a novel target-triggered "turn on" chemiluminescence (CL) sensor for alpha-fetoprotein (AFP) detection. Magnetic graphene oxide (MGO) was functionalized with the complementary sequence of the AFP aptamer (cDNA), and then MGO-cDNA was linked to aptamer modified luminol (Apt-luminol) through the complementary base pairing effect. The functionalized magnetic graphene oxide (MGO-cDNA/Apt-luminol) was prepared as a specific magnetic separation and signal switch material. ZnONPs-Au@CuMOFs shows excellent catalytic performance and was used as a catalyst for the luminol-H2O2 reaction. AFP will specifically recognize and bind to Apt on MGO-cDNA/Apt-luminol when AFP is present, which causes luminol release and triggers the CL reaction. The released luminol encounters ZnONPs-Au@CuMOFs and produces strong CL intensity. Therefore, a novel target-triggered "turn on" CL method with high selectivity and sensitivity for detecting AFP has been established. The linear range and detection limit were 1.0 × 10-4-50 ng mL-1 and 4.2 × 10-5 ng mL-1, respectively. The sensor also exhibited good selectivity, reproducibility and stability, and was finally used for AFP detection in serum samples.
More Related Videos
Related Concept Videos
Labeling DNA Probes
Radioisotopes, fluorophores, or small molecule binding partners like biotin or digoxigenin, are the most widely used reporter tags for labeling DNA probes. These labels can be attached to the probe DNA molecule via...
Photoluminescence: Applications