Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector
Xinming Zhang1,2, Hanwei Li1, Haowen Wang1
1Department of Clinical Engineering, Nanfang Hospital, Southern Medical University, Guangzhou, PR China.
Background:
Metal implants may affect the image quality, iodine concentration (IC), and CT Hounsfield unit (HU) quantification accuracy.
Purpose:
To investigate the quantitative accuracy of IC and HU from dual-layer spectral detector (DLCT) in the presence of metal artifacts.
Material And Methods:
An experimental cylindrical phantom containing eight iodine inserts and two metal inserts was designed. The phantom underwent scanning at three radiation dose levels and two tube voltage settings. A set of conventional images (CIs), virtual monoenergetic images (VMIs), and iodine concentration maps (ICMs) were generated and measured for all the eight iodine inserts. Quantitative indicators of mean absolute percentage error (MAPE), artifact index (AI), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and standard deviation (SD) on CIs and VMIs were calculated for IC and HU. Subjective score evaluation was also conducted.
Results:
The MAPEiodine values of all regions of interest across different scanning configurations were all <5%. Almost all APEiodine values were <5%, indicating that metal artifacts had little impact on IC measurements. When the tube voltage was fixed, the SD value of attenuation decreased with the increase of the tube current; this is also true when the tube current was fixed. The middle energy reconstructions seemed to give a good balance between reducing artifacts and improving contrast.
Conclusion:
VMIs from DLCT can reduce metal artifacts, the accuracy of IC quantification is not sensitive to imaging parameters. In summary, metal implants exhibit minimal impact on image quality and IC quantification accuracy in reconstructed images from DLCT.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...