Clinical and Analytical Performance of a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay
Luca Koechlin1, Jasper Boeddinghaus2, Pedro Lopez-Ayala3
1Cardiovascular Research Institute Basel, and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland; Department of Cardiac Surgery, University Hospital Basel, University of Basel, Basel, Switzerland; GREAT Network, Rome, Italy.
Background:
Point-of-care (POC) high-sensitivity cardiac troponin assays may further accelerate the diagnosis of myocardial infarction (MI).
Objectives:
This study sought to assess the clinical and analytical performance of the novel high-sensitivity cardiac troponin I (hs-cTnI)-SPINCHIP POC test.
Methods:
Adult patients presenting with acute chest discomfort to the emergency department were enrolled in an international, diagnostic, multicenter study. The final diagnosis was centrally adjudicated by 2 independent cardiologists using all clinical information. We compared the discriminatory performance of hs-cTnI-SPINCHIP with current established central laboratory assays and derived an assay-specific hs-cTnI-SPINCHIP 0/1-hour algorithm. Secondary analyses included sample type comparisons (whole blood, fresh/frozen plasma, and capillary finger prick) and precision analysis.
Results:
MI was the adjudicated final diagnosis in 214 (19%) of 1,102 patients. Area under the receiver-operating characteristic curve was 0.94 (95% CI: 0.92-0.95) for hs-cTnI-SPINCHIP vs 0.94 (95% CI: 0.92-0.95) for hs-cTnI-Architect (P = 0.907) and 0.93 (95% CI: 0.91-0.95) for high-sensitivity cardiac troponin T Elecsys (P = 0.305). A cutoff <7 ng/L at presentation (if chest pain onset was >3 hours) or <7 ng/L together with a 0/1-hour delta of <4 ng/L ruled out 51% with a sensitivity and negative predictive value of 100% (95% CI: 97.7%-100%) and 100% (95% CI: 99.0%-100%), respectively. A hs-cTnI-SPINCHIP concentration ≥36 ng/L or a 0/1-hour delta ≥11 ng/L ruled in 27% with a specificity and positive predictive value of 90.9% (95% CI: 88.3%-92.9%) and 72.9% (95% CI: 66.4%-78.6%), respectively. Bootstrap internal validation confirmed excellent diagnostic performance. High agreement was observed between different sample types.
Conclusions:
The SPINCHIP hs-cTnI POC test has very high diagnostic accuracy. Its assay-specific 0/1-hour algorithm achieved very high sensitivity/negative predictive value and specificity/positive predictive value for rule-out/in MI. (Advantageous Predictors of Acute Coronary Syndromes Evaluation [APACE] Study [APACE]; NCT00470587).
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
08:58Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...