Subtraction of single-photon emission computed tomography (SPECT) in radioembolization: a comparison of four methods
Camiel E M Kerckhaert1, Hugo W A M de Jong2, Marjolein B M Meddens2
1Radiology and Nuclear Medicine, Utrecht University and University Medical Center Utrecht, P.O. Box 85500, 3508 GA, Utrecht, Netherlands. c.e.m.kerckhaert-2@umcutrecht.nl.
Background:
Subtraction of single-photon emission computed tomography (SPECT) images has a number of clinical applications in e.g. foci localization in ictal/inter-ictal SPECT and defect detection in rest/stress cardiac SPECT. In this work, we investigated the technical performance of SPECT subtraction for the purpose of quantifying the effect of a vasoconstricting drug (angiotensin-II, or AT2) on the Tc-99m-MAA liver distribution in hepatic radioembolization using an innovative interventional hybrid C-arm scanner. Given that subtraction of SPECT images is challenging due to high noise levels and poor resolution, we compared four methods to obtain a difference image in terms of image quality and quantitative accuracy. These methods included (i) image subtraction: subtraction of independently reconstructed SPECT images, (ii) projection subtraction: reconstruction of a SPECT image from subtracted projections, (iii) projection addition: reconstruction by addition of projections as a background term during the iterative reconstruction, and (iv) image addition: simultaneous reconstruction of the difference image and the subtracted image.
Results:
Digital simulations (XCAT) and phantom studies (NEMA-IQ and anthropomorphic torso) showed that all four methods were able to generate difference images but their performance on specific metrics varied substantially. Image subtraction had the best quantitative performance (activity recovery coefficient) but had the worst visual quality (contrast-to-noise ratio) due to high noise levels. Projection subtraction showed a slightly better visual quality than image subtraction, but also a slightly worse quantitative accuracy. Projection addition had a substantial bias in its quantitative accuracy which increased with less counts in the projections. Image addition resulted in the best visual image quality but had a quantitative bias when the two images to subtract contained opposing features.
Conclusion:
All four investigated methods of SPECT subtraction demonstrated the capacity to generate a feasible difference image from two SPECT images. Image subtraction is recommended when the user is only interested in quantitative values, whereas image addition is recommended when the user requires the best visual image quality. Since quantitative accuracy is most important for the dosimetric investigation of AT2 in radioembolization, we recommend using the image subtraction method for this purpose.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...