Handheld dynamometry: Validity and reliability of measuring hip joint rate of torque development and peak torque
Katherine McNabb1, María B Sánchez1, James Selfe1
1Department of Health Professions, Faculty of Health and Education, Manchester Metropolitan University, Manchester, United Kingdom.
Introduction:
Measuring rate of torque development (RTD) and peak torque (PT) for hip muscle performance presents challenges in clinical practice. This study investigated the construct validity of a handheld dynamometer (HHD) versus an isokinetic dynamometer (IKD), and intra-rater repeated reliability for RTD and PT and their relationship in hip joint movements.
Methods:
Thirty healthy individuals (mean age = 30 ± 8 years, 13 males) underwent two test sessions in a single day. RTD (0-50, 0-100, 0-150, 0-200ms) and PT normalised to body mass in maximal voluntary isometric contractions were measured using a HHD and an IKD in hip flexion, extension, abduction, adduction, internal and external rotation.
Results:
For validity between the devices, RTD0-50 exhibited the largest significant systematic bias in all hip movements (3.41-11.99 Nm·s-1 kg-1) and widest limits-of-agreement, while RTD0-200 had the lowest bias (-1.33-3.99 Nm·s-1 kg-1) and narrowest limits-of-agreement. For PT, agreement between dynamometers was observed for hip flexion (0.08 Nm·kg-1), abduction (-0.09 Nm·kg-1), internal (-0.01 Nm·kg-1), and external rotation (0.05 Nm·kg-1). For reliability, intra-rater intraclass correlation coefficient (ICC2,1) ranged from moderate to good in RTD0-50 and RTD0-100 (0.5-0.88), and good to excellent in RTD0-150 and RTD0-200 (0.87-0.95) in all movements. The HHD displayed excellent intra-rater, relative reliability values (ICC2,1) in all movements (0.85-0.95). Pearson's correlation revealed good linear correlation between PT and RTD0-150 and RTD0-200 in all movements (r = .7 to .87, p = < .001).
Conclusion:
Validity analysis demonstrated significant systematic bias and lack of agreement in RTD measures between the HHD and IKD. However, the HHD displays excellent to moderate intra-rater, relative reliability for RTD and PT measures in hip movements. Clinicians may use the HHD for hip muscle PT assessment but note, late phase RTD measures are more reliable, valid, and relate to PT than early phase RTD. Additionally, the correlation between RTD and PT at various time epochs was examined to better understand the relationship between these measures.
More Related Videos
07:30Muscle Imbalances: Testing and Training Functional Eccentric Hamstring Strength in Athletic Populations
Published on: May 1, 2018
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Related Concept Videos
Upward Impending Motion
To better comprehend how a screw jack functions, consider the completely unraveled thread as a block in contact with the...
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used to...