Related Experiment Video
Updated: Jun 16, 2025

Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Assessment of blood pressure variability: characteristics and comparison of blood pressure measurement methods
Keisuke Narita1,2, Daichi Shimbo3, Kazuomi Kario4
1Columbia Hypertension Laboratory, Department of Medicine, Columbia University Irving Medical Center, New York, NY, USA. kn2676@cumc.columbia.edu.
Abstract:
Previous studies have reported that blood pressure variability (BPV) is associated with the risk of cardiovascular events independent of blood pressure (BP) levels. While there is little evidence from intervention trials examining whether suppressing BPV is useful in preventing cardiovascular disease, it is suggested that detection of abnormally elevated BPV may be useful in reducing cardiovascular events adding by complementing management of appropriate BP levels. Cuffless BP devices can assess beat-to-beat BPV. Although cuffless BP monitoring devices have measurement accuracy issues that need to be resolved, this is an area of research where the evidence is accumulating rapidly, with many publications on beat-to-beat BPV over several decades. Ambulatory BP monitoring (ABPM) can assess 24-hour BPV and nocturnal dipping patterns. Day-to-day BPV and visit-to-visit BPV are assessed by self-measured BP monitoring at home and office BP measurement, respectively. 24 h, day-to-day, and visit-to-visit BPV have been reported to be associated with cardiovascular prognosis. Although there have been several studies comparing whether ABPM and self-measured BP monitoring at home is the superior measurement method of BPV, no strong evidence has been accumulated that indicates whether ABPM or self-measured home BP is superior. ABPM and self-measured BP monitoring have their own advantages and complement each other in the assessment of BPV.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring