Related Experiment Video
Updated: Jun 16, 2025

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Complete pacemaker failure following lightning strike injury: A case report
Gustavo Gomes Torres1, William Santos de Oliveira1, Nestor Rodrigues de Oliveira Neto1
1Department of Cardiology, Hospital Universitário Onofre Lopes, Universidade Federal do Rio Grande do Norte, Natal, Brazil.
Introduction:
Modern cardiovascular implantable electronic devices (CIEDs) have mechanisms that prevent damage from external electric shocks, and malfunction following accidental electrocution is rare. However, the effects of lightning injuries in patients with CIEDs are uncertain.
Case Presentation:
A 74-year-old man with a dual-chamber pacemaker due to complete heart block was struck by a lightning while farming. He had no serious injury at the time and sought medical evaluation 1 month later, when he presented with asymptomatic bradycardia. Device interrogation suggested major battery and lead damage, requiring extraction and subsequent placement of a new pacing system.
Discussion:
While a previous report depicted pacing threshold elevation without extensive device impairment, our patient presented with major damage to the whole pacing system. The factors contributing to these divergent outcomes are unclear. Differences in injury mechanism, pacemaker model, and the pattern of electric current dispersion within the device may each play a part in this discrepancy.
More Related Videos
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias