PETcoil: first results from a second-generation RF-penetrable TOF-PET brain insert for simultaneous PET/MRI
Qian Dong1, Muhammad Nasir Ullah1, Derek Innes1
1Molecular Imaging Instrumentation Laboratory, Department of Radiology, Stanford University, Stanford, CA, United States of America.
Physics in Medicine and Biology
|August 21, 2024
Summary
This study introduces a new MR-compatible PET insert with improved spatial resolution and sensitivity. Initial performance tests show stable operation and enhanced image quality with time-of-flight reconstruction.
Area of Science:
- Medical Imaging
- Physics
- Engineering
Background:
- Simultaneous Positron Emission Tomography (PET)/Magnetic Resonance Imaging (MRI) offers integrated anatomical, functional, and molecular insights for disease diagnosis.
- Advancements in PET/MRI technology are crucial for improving diagnostic capabilities.
Purpose of the Study:
- To report the initial system performance of a novel, second-generation, MR-compatible, radiofrequency-penetrable Time-of-Flight PET insert.
- To evaluate the spatial resolution, sensitivity, and stability of the full-ring PET insert.
Main Methods:
- Developed a full-ring PET insert with smaller scintillation crystals and ring diameter for enhanced resolution and sensitivity.
- Utilized 12,288 LYSO-SiPM channels with a crystal size of 3.2 × 3.2 × 20 mm³.
- Conducted stability tests over ten hours and resolution phantom studies (2.8 mm rods).
- Performed initial MR compatibility studies with the unpowered PET ring.
Main Results:
- Preserved global photopeak energy resolution (11.74 ± 0.03 % FWHM) and coincidence time resolution (238.1 ± 0.5 ps FWHM).
- Achieved successful resolution of 2.8 mm rods in phantom studies, with VPR of 0.28 ± 0.08 (no TOF) and 0.24 ± 0.07 (TOF).
- Demonstrated stable system performance (<1% change over 10 hours) and enhanced image quality with TOF.
- No observable MR artifacts detected during initial unpowered MR compatibility tests, though SNR degradation (~30%) occurred with the body coil.
Conclusions:
- The developed MR-compatible TOF-PET insert demonstrates promising performance with high spatial resolution and sensitivity.
- The system exhibits stable operation and potential for improved image quality through TOF reconstruction.
- Further development is needed for a dedicated receive-only MR coil to mitigate SNR degradation.
Related Concept Videos
Positron Emission Tomography
4.1K
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
4.1K
Brain Imaging
219
Brain imaging technologies provide critical insights into both the structure and function of the human brain, enabling medical professionals and researchers to diagnose, study, and treat neurological disorders or psychiatric disorders more effectively.
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
219