Related Experiment Video
Updated: Jun 15, 2025

Author Spotlight: Exploring Venous Waveforms for Non-Invasive Respiratory Monitoring in Pigs
Published on: March 8, 2024
Racial Disparities in Pulse Oximetry, in COVID-19 and ICU Settings
Carmen A T Reep1, Lucas M Fleuren1, Leo Heunks1,2
1Department of Intensive Care, Erasmus Medical Center, Rotterdam, the Netherlands.
Objectives Background:
This study aimed to assess the impact of race on pulse oximetry reliability, taking into account Spo2 ranges, COVID-19 diagnosis, and ICU admission.
Design:
Retrospective cohort study covering admissions from January 2020 to April 2024.
Setting:
National COVID Cohort Collaborative (N3C) database, consisting of electronic health records from 80 U.S. institutions.
Patients/Subjects:
Patients were selected from the N3C database based on the availability of data on self-identified race and both pulse oximetry estimated Spo2 and Sao2. Subgroups included patients in ICU and non-ICU settings, with or without a diagnosis of COVID-19 disease.
Interventions:
None.
Measurements And Main Results:
The agreement between Spo2 and Sao2 was assessed across racial groups (American Indian or Alaska Native, Asian, Black, Hispanic or Latino, Pacific Islander, and White). Each patient's initial Sao2 measurement was matched with the closest Spo2 values recorded within the preceding 10-minute time frame. The risk of hidden hypoxemia (Spo2 ≥ 88% but Sao2 < 88%) was determined for various Spo2 ranges, races, and clinical scenarios. We used a generalized logistic mixed-effects model to evaluate the impact of relevant variables, such as COVID-19, ICU admission, age, sex, race, and Spo2, on the risk of hidden hypoxemia, while accounting for the random effects within each hospital. A total of 80,541 patients were included, consisting of 596 American Indian or Alaska Native, 2,729 Asian, 11,889 Black, 13,154 Hispanic or Latino, 221 Pacific Islander, and 51,952 White individuals. Discrepancies between Spo2 and Sao2 were observed across all racial groups, with the most pronounced bias in Black patients. Hidden hypoxemia rates were higher in Black patients across all Spo2 subgroups, for all clinical scenarios. The odds of hidden hypoxemia were higher for Black and Hispanic or Latino patients and for those with COVID-19 disease.
Conclusions:
Race significantly impacts pulse oximetry reliability. Not only Black and Hispanic or Latino patients were at higher risk for hidden hypoxemia, but also those admitted with a COVID-19 diagnosis. Future in-depth explorations into the underlying causes and potential solutions are needed.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...