D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case

Marija Milić1,2, Dejana Brkić Barbarić3, Iva Lukić1,2

  • 1Clinical Institute of Laboratory Diagnostics, University Hospital Centre Osijek, Osijek, Croatia.

Biochemia Medica
|August 22, 2024
PubMed
Summary

Heterophilic antibodies can interfere with D-dimer testing, leading to inaccurate results. This case report details how SARS-CoV-2 infection may cause such interference, necessitating careful laboratory investigation and clinical-laboratory collaboration.

Related Concept Videos

Atomic Absorption Spectroscopy: Interference01:25

Atomic Absorption Spectroscopy: Interference

Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
704
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
419
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
6.0K