Related Experiment Video
Updated: Jun 15, 2025

Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model
Published on: May 24, 2024
Pedicle screw placement in the cervical vertebrae using augmented reality-head mounted displays: a cadaveric
Miguel A Ruiz-Cardozo1, Karma Barot1, Samuel Brehm1
1Department of Neurosurgery, Washington University School of Medicine, St. Louis, MO, USA.
Background:
The accurate and safe positioning of cervical pedicle screws is crucial. While augmented reality (AR) use in spine surgery has previously demonstrated clinical utility in the thoracolumbar spine, its technical feasibility in the cervical spine remains less explored.
Purpose:
The objective of this study was to assess the precision and safety of AR-assisted pedicle screw placement in the cervical spine.
Study Design:
In this experimental study, 5 cadaveric cervical spine models were instrumented from C3 to C7 by 5 different spine surgeons. The navigation accuracy and clinical screw accuracy were evaluated.
Methods:
Postprocedural CT scans were evaluated for clinical accuracy by 2 independent neuroradiologists using the Gertzbein-Robbins scale. Technical precision was assessed by calculating the angular trajectory (°) and linear screw tip (mm) deviations in the axial and sagittal planes from the virtual pedicle screw position as recorded by the AR-guided platform during the procedure compared to the actual pedicle screw position derived from postprocedural imaging.
Results:
A total of forty-one pedicle screws were placed in 5 cervical cadavers, with each of the 5 surgeons navigating at least 7 screws. Gertzbein-Robbins grade of A or B was achieved in 100% of cases. The mean values for tip and trajectory errors in the axial and sagittal planes between the virtual versus actual position of the screws was less than 3 mm and 30°, respectively (p<.05). None of the cervical screws violated the cortex by more than 2 mm or displaced neurovascular structures.
Conclusions:
AR-assisted cervical pedicle screw placement in cadavers demonstrated clinical accuracy comparable to existing literature values for image-guided navigation methods for the cervical spine.
Clinical Significance:
This study provides technical and clinical accuracy data that supports clinical trialing of AR-assisted subaxial cervical pedicle screw placement.