A dosimetric comparison of brachytherapy sources for endometrial cancer: an electronic brachytherapy and an iridium-192 source with multichannel cylinders and a three-dimensional technique
Related Experiment Video
Updated: Aug 8, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Wenhui Wang1, Bei Wang1, Lang Yu2
1Department of Radiotherapy, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, No. 1 Shuaifuyuan Wangfujing Dongcheng District, Beijing, 100730, China.
Background And Purpose:
Ir192 vaginal brachytherapy (IBT) is commonly used for patients with postoperative endometrial cancer (EC). We devised a novel multichannel vaginal applicator that could be equipped with an electronic brachytherapy (EBT) device. We aimed to explore the differences in physical parameters between the EBT and IBT.
Materials And Methods:
This retrospective study included 20 EC patients who received adjuvant IBT from March 1, 2023, to May 1, 2023. Multichannel vaginal cylinders were used, and three-dimensional plans were generated. We designed an electronic multichannel vaginal applicator model and simulated a three-dimensional EBT plan. In order to ensure comparability, D90 of the CTV for the EBT plan was normalized to be equivalent to that of the IBT plan for the same patient.
Results:
Twenty EBT plans were compared with 20 IBT plans. Results showed, the mean D90 value of clinical target volume (CTV) was 536.1 cGy for both treatment plans. For the mean dose of CTV, the EBT was significantly greater (738.3 vs. 684.3 cGy, p = 0.000). There was no significant difference in CTV coverage between the EBT and IBT plans. For high-dose areas (V200% and V150%), the EBTs were significantly greater. There were no significant differences in the maximum doses to the vaginal mucosa between the EBT and IBT, whether at the apex or in the middle segment. For the bladder and rectum, both the low-dose area and high-dose area were significantly lower in the EBT plans. For the conformity index, there was no significant difference between the EBT and IBT plans. For the dose homogeneity index, the EBT value was lower.
Conclusion:
In conclusion, under the premise of a three-dimensional brachytherapy plan, for patients receiving multichannel vaginal applicator brachytherapy, compared with IBT, EBT could reduce the dose to the surrounding organs at risk while maintaining the dose in the target area.
More Related Videos
Related Concept Videos
Cancer Therapies
However, cancer treatments can pose several challenges, as therapies used to kill cancer cells are generally also toxic to normal cells. Moreover, cancer cells mutate rapidly and can develop resistance to chemical agents or radiation therapy. Besides, all types of cancer cells may not respond to the same therapy. Some cancer cells respond to one...