Automated devices for identifying peripheral arterial disease in people with leg ulceration: an evidence synthesis
Dwayne Boyers1, Moira Cruickshank2, Lorna Aucott2
1Health Economics Research Unit, University of Aberdeen, Aberdeen, UK.
Background:
Peripheral artery disease is a common condition caused by narrowing/blockage of the arteries, resulting in reduced blood supply. Peripheral artery disease is associated with an increased risk of vascular complications, but early treatment reduces mortality and morbidity. Leg ulcers are long-lasting wounds, usually treated by compression therapy. Compression therapy is not suitable for people with peripheral artery disease, as it can affect the arterial blood supply. In clinical practice, people with peripheral artery disease are identified by measurement of the ankle-brachial pressure index using a sphygmomanometer and manual Doppler device. However, this method can be uncomfortable for people with leg ulcers and automated devices have been proposed as a more acceptable alternative. The objective of this appraisal was to summarise the clinical and cost-effectiveness evidence on the use of automated devices to detect peripheral artery disease in people with leg ulcers.
Methods:
.
Clinical Effectiveness:
To identify reports of relevant studies, we searched major electronic databases and scrutinised the information supplied by the manufacturers of the automated devices under investigation. Due to the lack of evidence on people with leg ulcers, we considered evidence from studies of any design assessing automated devices versus an acceptable reference device in any population receiving ankle-brachial pressure index assessment. We summarised information on diagnostic accuracy of the automated devices and level of agreement with the reference device. For each device, when data permit, we pooled data across studies by conducting random-effects meta-analyses using a Hierarchical Summary Receiving Operating Characteristics model.
Cost-Effectiveness:
An economic model comprising a decision tree (24 weeks) and Markov models to capture lifetime costs and quality-adjusted life-years associated with venous, arterial and mixed aetiology disease in leg ulcer patients. Analyses were conducted from a United Kingdom National Health Service and Personal Social Services perspective. Costs and quality-adjusted life-years were discounted at 3.5% per year. Deterministic and several probabilistic analyses were used to capture uncertainty surrounding a range of optimistic and pessimistic assumptions about the impact of automated tests on health outcomes (ulcer healing and requirement for invasive management of arterial disease).
Results:
.
Clinical Effectiveness:
From the 116 records retrieved by the electronic searches, we included 24 studies evaluating five devices (BlueDop Vascular Expert, BOSO ABI-System 100, Dopplex Ability, MESI ankle-brachial pressure index MD and WatchBP Office ABI). Two studies assessing people with leg ulcers found that automated devices often gave higher ankle-brachial pressure index readings than manual Doppler (underestimation of arterial disease). In the 22 studies involving people without leg ulcers, automated devices generally demonstrated good specificity and moderate specificity. Meta-analysis of 12 studies showed a pooled sensitivity of 64% (95% confidence interval 57% to 71%) and a pooled specificity of 96% (95% confidence interval 92% to 98%) for detection of peripheral artery disease.
Cost-Effectiveness:
Automated devices cost less than manual Doppler to deliver. However, increased risks of invasive treatment requirements for inappropriately compressed arterial/mixed ulcers due to false-negative results, and increased healing times due to delayed compression of false-positive test results mean that in most scenarios manual Doppler was less costly and had slightly higher quality-adjusted life-years than automated devices. Results are highly uncertain, dependent on many assumptions and should be interpreted cautiously.
Limitations And Conclusions:
The limited evidence identified for each automated device, especially in people with leg ulcers, and its clinical heterogeneity precludes any firm conclusions on the diagnostic performance and cost-effectiveness of these devices in clinical practice.
Study Registration:
This study is registered as PROSPERO CRD42022327588.
Funding:
This award was funded by the National Institute for Health and Care Research (NIHR) Evidence Synthesis programme (NIHR award ref: NIHR135478) and is published in full in Health Technology Assessment; Vol. 28, No. 37. See the NIHR Funding and Awards website for further award information.
More Related Videos
07:25Predicting Amputation using Local Circulating Mononuclear Progenitor Cells in Angioplasty-treated Patients with Critical Limb Ischemia
Published on: September 22, 2020
04:34A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Related Concept Videos
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Pulse Assessment Sites
Assessing Blood pressure in the Leg
Preparation:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...