Response characterization of radiochromic OC-1 films in photon, electron, and proton beams
Qinghao Chen1, Xiandong Zhao1, Jufri Setianegara1
1Department of Radiation Oncology, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
Background:
Radiochromic film (RCF) dosimeters with their high spatial resolution and tissue equivalent properties are conveniently used for two-dimensional and small-field dosimetry. OC-1 is a new model of RCF dosimeter that was commercially introduced recently. Due to its novelty there is a need to characterize its response in various radiation beam types.
Purpose:
To study the response of OC-1 RCFs to megavoltage clinical x-ray, electron, and proton beams, as well as kilovoltage x-ray beams used in a small animal research irradiator.
Materials And Methods:
OC-1 RCFs were cut into ∼4 × 4 cm2 pieces. RCF samples were irradiated at various dose levels in the range 0.5-120 Gy using different modalities; a small animal radiation research platform (SARRP) (220 kVp), a medical linear accelerator (6 MV, 10 MV, 15 MV, 6 MV FFF, 10 MV FFF photon beams, as well as 6 and 20 MeV electron beams), and a gantry-mounted proton therapy synchrocyclotron. In order to study any dependency on the fractionation scheme, same dose was delivered at several fractions to a set of films. Different dose rates in the range 200-600 MU/min were delivered to a set of films to investigate any dose rate dependency. The films were scanned pre-irradiation and at 48 h post-irradiation using a flatbed scanner. The net optical density (OD) was measured for red, green, and blue color channel for each film. The orientation dependency was studied by scanning the films at eight different orientations. In order to study the temporal evolution of the response of the films, film samples were irradiated at 10 and 50 Gy using 6 MV photon beams and were scanned upon irradiation at certain time intervals up to 3 months. The spectral response of the films were studied over the visible range using a spectrometer.
Results:
For megavoltage photon, electron, and plateau region of the proton beams, we did not observe a significant dependency on the beam quality, dose rate, and fractionation scheme. At the kV beam, an unusual over-response was observed in the films' net OD. An orientation dependency in the response of the films with a sinusoidal trend was observed. The response of the films increased with time following a double or triple exponential trend. The spectral absorption peaks were blue-shifted with dose.
Conclusion:
OC-1 RCFs were found to be reliable dosimeters with no significant energy dependency in MV range for photon and electron beams including the FFF beams. They over-respond when irradiated by kV x-ray beams compared to MV x-ray beams. Caution must be exercised to maintain the orientation of the films when scanning. Due to the temporal growth in the net OD of the films, same post-irradiation time interval must be used for scanning the calibration and test films.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
The Electromagnetic Spectrum
X-ray Imaging
Interaction of EM Radiation with Matter: Spectroscopy
Isotopes and Radioisotopes
An isotope containing...