Ventricular tachycardia ablation with pentaspline pulsed field technology in two patients with ischemic
Gaetano Fassini1, Elio Zito1, Lorenzo Bianchini1
1Department of Clinical Electrophysiology & Cardiac Pacing, Centro Cardiologico Monzino, IRCCS, Milan, Italy.
Introduction:
Due to its unique features, pulsed field ablation (PFA) could potentially overcome some limitations of current radiofrequency (RF) ventricular tachycardia (VT) ablation. However, data on the use of PFA in this setting are currently scarce.
Methods:
Two patients with ischemic cardiomyopathy and previously failed RF VT ablations were treated with PFA.
Results:
A total of 18 bipolar applications (case1) and seven bipolar applications (case2) were delivered to the infero-lateral and infero-septal areas (case1) and to the apical lateral left ventricular (LV) wall (case2), placing the catheter adjacent to the LV wall in the flower configuration. A rapid cessation of VT and restoration of sinus rhythm were observed during PFA delivery in both cases. Further applications were delivered to achieve complete elimination of late potentials. In case 1, during the in-hospital stay, ECG monitoring did not show VT recurrences. Six-month follow-up was uneventful, with no VT recurrences at ICD interrogation. In case 2, due to postdischarge VT recurrences, a second RF procedure was scheduled 1 month later. The voltage map performed in sinus rhythm showed a low-voltage zone located at the anterolateral wall, near the previous ablation site. Numerous late potentials were recorded. At the 6-month follow-up, no further VT recurrences were documented after RF redo ablation.
Conclusion:
While the speed of application and potential transmural effect can facilitate the ablation of large diseased endocardial areas, early loss of contact due to difficult pentaspline catheter manipulation in the LV could lead to insufficient contact force and, consequently, inadequate energy penetration.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care