Simultaneous Multiparameter Mapping of the Liver in a Single Breath-Hold or Respiratory-Triggered Acquisition Using
Mary Kate Manhard1,2, Anandh Kilpattu Ramaniharan1, Jean A Tkach1,2
1Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
Journal of Magnetic Resonance Imaging : JMRI
|August 27, 2024
Summary
The multi-inversion spin and gradient echo (MI-SAGE) technique provides reliable liver T1, T2, and T2* measurements comparable to conventional methods. This advanced technique offers strong repeatability and reproducibility for chronic liver disease assessment.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Medical Physics
- Radiology
Background:
- Quantitative parametric mapping is crucial for noninvasive assessment of chronic liver disease.
- Conventional techniques require multiple breath-holds and have limited coverage.
- Novel methods are needed to improve efficiency and coverage.
Purpose of the Study:
- To evaluate a multi-inversion spin and gradient echo (MI-SAGE) technique.
- To assess its capability for simultaneous estimation of liver T1, T2, and T2*.
- To compare MI-SAGE with conventional parametric mapping sequences.
Main Methods:
- Prospective study involving 16 participants (adult and pediatric) with and without liver disease.
- Utilized 1.5T MRI with single breath-hold and respiratory-triggered MI-SAGE sequences.
- Compared MI-SAGE with breath-hold MOLLI (T1), GRASE (T2), and mGRE (T2*) sequences.
- Statistical analysis included Bland-Altman analysis and intraclass correlation coefficients (ICC).
Main Results:
- MI-SAGE T1, T2, and T2* estimates showed good agreement with conventional methods, with minor biases.
- Mean biases for T1, T2, T2* were (72, -3, 2) msec for breath-hold and (36, -3, 4) msec for respiratory-triggered MI-SAGE.
- All MI-SAGE estimates demonstrated strong repeatability and reproducibility (ICC > 0.72).
Conclusions:
- The MI-SAGE technique yields hepatic T1, T2, and T2* estimates comparable to conventional MRI methods.
- A slight underestimation of T1 values was observed compared to MOLLI.
- Both breath-hold and respiratory-triggered MI-SAGE parameter maps exhibit robust repeatability and reproducibility.