Related Experiment Video For computed radiography
Updated: Jun 15, 2025

Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Influence of Examiner Experience on the Measurement of Bone-Loss by Low-Dose Cone-Beam Computed Tomography: An Ex
Maurice Ruetters1, Korallia Alexandrou2, Antonio Ciardo1
1Department of Operative Dentistry, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany.
Abstract:
The aim of this study was to investigate the influence of examiner experience on measurements of bone-loss using high-dose (HD) and low-dose (LD) CBCT. Three diagnosticians with varying levels of CBCT interpretation experience measured bone-loss from CBCT scans of three cadaveric heads at 30 sites, conducting measurements twice. Between the first and second measurements, diagnostician 2 and diagnostician 3 received training in LD-CBCT diagnostics. The diagnosticians also classified the certainty of their measurements using a three-grade scale. The accuracy of bone-loss measurements was assessed using the absolute difference between observed and clinical measurements and compared among diagnosticians with different experience levels for both HD and LD-CBCT. At baseline, there was a significant difference in measurement accuracy between diagnostician 1 and diagnostician 2, and between diagnostician 1 and diagnostician 3, but not between diagnostician 2 and diagnostician 3. Training improved the accuracy of both HD-CBCT and LD-CBCT measurements in diagnostician 2, and of LD-CBCT measurements in diagnostician 3. Regarding measurement certainty, there was a significant difference among diagnosticians before training. Training enhanced the certainty for diagnosticians 2 and 3, with a significant improvement noted only for diagnostician 3. Examiner experience level significantly impacts the accuracy and certainty of bone-loss measurements using HD- and LD-CBCT.
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...

