Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study
Maria Francesca Sfondrini1, Maurizio Pascadopoli1, Paola Gandini1
1Unit of Orthodontics and Pediatric Dentistry, Section of Dentistry, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy.
Stainless steel multibraided fixed retainers are safe for MRI scans. Testing showed no significant heating or bonding issues at 1.5-tesla (T) and 3-T magnetic resonance imaging (MRI) examinations.
Area of Science:
- Dental materials science
- Biomaterials engineering
- Medical imaging physics
Background:
- Orthodontic fixed retainers are commonly used to maintain tooth alignment post-treatment.
- Patients with fixed retainers may require magnetic resonance imaging (MRI) for diagnostic purposes.
- Concerns exist regarding potential heating and compromised bonding of metallic orthodontic appliances during MRI.
Purpose of the Study:
- To evaluate the in vitro heating effects and bonding efficacy of stainless steel multibraided fixed retainers under 1.5-tesla (T) and 3-T MRI conditions.
- To determine if different wire diameters of the retainers influence heating or bonding strength after MRI exposure.
Main Methods:
- 180 human mandibular incisors were used to create 45 four-tooth specimens.
- Handmade multibraided fixed retainers (0.008″, 0.010″, 0.012″ wire diameters) were bonded to the incisors.
- Specimens were divided into nine groups: three for 1.5 T MRI, three for 3 T MRI, and three controls (no MRI).
- Temperature changes, shear bond strength (SBS), and adhesive remnant index (ARI) were measured before and after MRI exposure.
Main Results:
- No significant temperature increase was observed after 1.5 T MRI exposure (p > 0.05).
- A significant temperature increase was noted in all 3 T MRI groups (p < 0.05), but changes were clinically irrelevant (< 1 °C).
- No significant differences in temperature variations were found across different wire diameters (p > 0.05).
Conclusions:
- No significant changes in shear bond strength (SBS) or adhesive remnant index (ARI) were observed after MRI exposure (p > 0.05).
- The tested stainless steel multibraided fixed retainers demonstrated negligible overheating and maintained bonding integrity.
- These findings indicate that the tested fixed retainers are safe for patients undergoing 1.5 T and 3 T MRI examinations.