Engaging vs. Non-Engaging Abutments: An In Vitro Study Evaluating Changes in Microgap and Screw Morphology
Fawaz M Alzoubi1, Mohammad Y Sabti1, Esra Alsarraf2
1Department of General Dental Practice, College of Dentistry, Kuwait University, P.O. Box 24923, Safat 13110, Kuwait.
This study found no significant difference in microgap size or screw damage between engaging and non-engaging abutments after cyclic loading in dental implants. These findings suggest similar performance regardless of abutment engagement type.
Area of Science:
- Dental Implantology
- Biomaterials Science
- Prosthodontics
Background:
- Microgaps in dental implants can harbor bacteria, potentially leading to peri-implantitis.
- Abutment design, specifically engaging (E) versus non-engaging (NE), may influence microgap formation and screw stability.
Purpose of the Study:
- To compare microgap size between E and NE abutments in single units and fixed partial dentures (FPDs).
- To evaluate screw morphology changes after cyclic loading (CL) in E and NE abutment configurations.
- To assess the influence of prosthetic design on microgap and screw integrity.
Main Methods:
- Thirty-six implants were divided into four groups: single units with E abutments, single units with NE abutments, three-unit FPDs with hemi-engaging abutments, and three-unit FPDs with two NE abutments.
- Microgap size was measured using a stereomicroscope.
- Screw morphology was qualitatively assessed using scanning electron microscopy (SEM) before and after cyclic loading (1.0 × 10^6 cycles) under axial and lateral (30°) forces.
Main Results:
- No significant differences in microgap size were observed between E and NE abutment groups, nor between different prosthetic designs.
- Cyclic loading did not significantly alter microgap sizes in any group.
- SEM revealed screw damage after CL, but the pattern of damage was similar between E and NE abutments.
Conclusions:
- Engaging and non-engaging abutments exhibit comparable microgap sizes in both single-unit and FPD configurations.
- Prosthetic design does not significantly impact microgap size.
- Screw morphology changes after cyclic loading are similar for both engaging and non-engaging abutments, indicating no distinct advantage of one over the other in terms of screw integrity under these conditions.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Studying the Cytoskeleton
Microcracking in Concrete