An Introduction to Ventra: A Programmable Abdominal Phantom for Training, Educational, Research, and Development
Salar Tayebi1, Robert Wise2,3,4, Ashkan Zarghami1
1Department of Electronics and Informatics, Vrije Universiteit Brussel (VUB), 1050 Brussels, Belgium.
Background:
Intra-abdominal pressure (IAP) is a critical parameter in the care of critically ill patients, as elevated IAP can lead to reduced cardiac output and organ perfusion, potentially resulting in multiple organ dysfunction and failure. The current gold standard for measuring IAP is an indirect technique via the bladder. According to the Abdominal Compartment Society's Guidelines, new measurement methods/devices for IAP must be validated against the gold standard.
Objectives:
This study introduces Ventra, an abdominal phantom designed to simulate different IAP levels, abdominal compliance, respiration-related IAP variations, and bladder dynamics. Ventra aims to facilitate the development and validation of new IAP measurement devices while reducing reliance on animal and cadaveric studies. Additionally, it offers potential applications in training and education for biomedical engineering students. This study provides a thorough explanation on the phantom's design and fabrication, which provides a low-cost solution for advancing IAP measurement research and education. The design concept, technical aspects, and a series of validation experiments determining whether Ventra is a suitable tool for future research are presented in this study.
Methods:
Ventra's performance was evaluated through a series of validation tests using a pressure gauge and two intra-gastric (Spiegelberg and CiMON) and two intra-bladder (Accuryn and TraumaGuard) pressure measurement devices. The mean and standard deviation of IAP recordings by each device were investigated. Bland-Altman analysis was used to evaluate bias, precision, limits of agreement, and percentage error for each system. Concordance analysis was performed to assess the ability of Ventra in tracking IAP changes.
Results:
The phantom demonstrated excellent agreement with reference pressure measurements, showing an average bias of 0.11 ± 0.49 mmHg. A concordance coefficient of 100% was observed for the phantom as well. Ventra accurately simulated different abdominal compliances, with higher IAP values resulting in lower compliance. Abdominal volume changes showed a bias of 0.08 ± 0.07 L/min, and bladder fill volume measurements showed an average difference of 0.90 ± 4.33 mL for volumes ranging from 50 to 500 mL.
Conclusion:
The validation results were in agreement with the research guidelines of the world abdominal society. Ventra is a reliable tool that will facilitate the development and validation of new IAP measurement devices. It is an effective educational tool for biomedical engineering students as well.
More Related Videos
Related Concept Videos
Vision
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Veins of the Abdomen and Pelvis
The inferior vena cava is fed by numerous smaller veins. The lumbar veins, for instance, drain the posterior abdominal wall, emptying both directly into the inferior vena cava and into the...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...