Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A
Jelena M Mihailovic1, Michael R Bruesewitz1, Joseph R Swicklik1
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Purpose:
A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA.
Methods:
A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images.
Results:
When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images.
Conclusion:
PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
More Related Videos
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...