Related Experiment Video
Updated: Jun 14, 2025

Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Mitigating Alarm Fatigue and Improving the Bedside Experience by Reducing Nonactionable Alarms
Jeffrey K Yang1, Felice Su1, Anna Graber-Naidich2
1Department of Pediatrics, Stanford University, Stanford, CA.
Implementing conditional bedside alarm triggers significantly reduced nonactionable alarms in cardiac and pediatric intensive care units. This quality improvement initiative enhanced nursing experience without compromising patient safety.
Area of Science:
- Critical Care Medicine
- Biomedical Engineering
- Patient Safety
Background:
- Alarms from bedside monitors are crucial for patient safety but often lead to alarm fatigue due to high frequency of nonactionable alerts.
- Reducing nonactionable alarms is essential to improve the effectiveness of clinical alarm systems and prevent adverse events.
Purpose of the Study:
- To evaluate the impact of conditional bedside alarm triggers on the frequency of nonactionable alarms.
- To assess whether these changes affect patient safety and satisfaction among nurses and families.
Main Methods:
- A single-center quality improvement initiative was conducted in an acute care cardiac unit and a pediatric intensive care unit.
- Bedside monitors were reprogrammed with hierarchical time delay and conditional alarm triggers after a baseline period.
- Alarm data and patient safety events were collected, and nurses and families were surveyed pre- and post-intervention.
Main Results:
- A significant reduction in median alarms per monitored patient day was observed: 75% in the pediatric intensive care unit and 82% in the acute care cardiac unit.
- Alarm reduction was sustained at a 2-year follow-up, with no increase in rapid response calls, emergent transfers, or code events.
- Nurses reported improved capacity to respond to alarms and fewer perceived nonactionable alarms; however, family surveys did not show improved sleep quality.
Conclusions:
- Modifications to bedside monitor alarm systems using conditional triggers effectively decreased total alarm frequency in critical care settings.
- These changes improved the care provider experience and maintained patient safety.
- Further research may be needed to address family satisfaction regarding sleep quality.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Heart Failure Drugs: Inhibitors of Renin-Angiotensin System
Nursing Diagnosis
The nursing diagnosis focuses on evidence-based...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.